Table 2. NI 601xE Analog Input Accuracy Specifications
Absolute Accuracy Relative Accuracy
Nominal % of Reading Offset Noise + Quantization (µV) Temp Absolute Accuracy Resolution (µV)
Range (V) 24 Hrs 90 Days 1 Year (
µ
V) Single Pt. Averaged Drift (%/°C) at Full Scale (mV) Single Pt. Averaged
±10.0 0.0058 0.0078 0.0100 397.2 526.4 45.8 0.0002 1.443 602.7 60.3
±5.0 0.0208 0.0228 0.0250 200.6 263.2 22.9 0.0007 1.474 301.4 30.1
±1.0 0.0208 0.0228 0.0250 43.3 52.6 4.6 0.0007 0.298 60.3 6.0
±0.1 0.0408 0.0428 0.0450 7.9 8.4 0.7 0.0012 0.054 9.6 1.0
0 to 10 0.0058 0.0078 0.0100 244.6 263.2 22.9 0.0002 1.268 301.4 30.1
0 to 5 0.0208 0.0228 0.0250 124.3 131.6 11.4 0.0007 1.386 150.7 15.1
0 to 1 0.0208 0.0228 0.0250 28.1 26.3 2.3 0.0007 0.280 30.1 3.0
0 to 0.1 0.0408 0.0428 0.0450 6.4 7.0 0.6 0.0012 0.052 8.4 0.8
±10.0 0.0075 0.0095 0.0117 815.4 1029.1 91.6 0.0005 2.077 1205.4 120.5
±5.0 0.0225 0.0245 0.0267 409.7 514.6 45.8 0.0010 1.791 602.7 60.3
±1.0 0.0225 0.0245 0.0267 85.1 102.9 9.2 0.0010 0.361 120.5 12.1
±0.1 0.0425 0.0445 0.0467 12.1 12.2 1.1 0.0015 0.060 14.5 1.4
0 to 10 0.0075 0.0095 0.0117 591.2 514.6 45.8 0.0005 1.807 602.7 60.3
0 to 5 0.0225 0.0245 0.0267 297.6 257.3 22.9 0.0010 1.656 301.4 30.1
0 to 1 0.0225 0.0245 0.0267 62.7 51.5 4.6 0.0010 0.334 60.3 6.0
0 to 0.1 0.0425 0.0445 0.0467 9.9 8.0 0.7 0.0015 0.057 9.6 1.0
Note: Accuracies are valid for measurements following an internal E Series Calibration. Averaged numbers assume dithering and averaging of 100 single-channel readings. Measurement accuracies are listed for
operational temperatures within ±1 °C of internal calibration temperature and ±10 °C of external or factory-calibration temperature. One-year calibration interval recommended. The Absolute Accuracy at Full Scale
calculations were performed for a maximum range input voltage (for example, 10 V for the ±10 V range) after one year, assuming 100 pt averaging of data. See overview on page 234 for example calculations.
PCI, AT, and DAQPadDAQCard
Absolute Accuracy
Nominal Range (V) % of Reading Temp Absolute Accuracy
Positive FS Negative FS 24 Hrs 90 Days 1 Year
Offset
(mV) Drift (%/°C) at Full Scale (mV)
10 -10 0.014 0.016 0.018 5.408 0.0002 7.208
10 0 0.014 0.016 0.018 2.966 0.0002 4.766
Note: Temp Drift applies only if ambient is greater than ±10 °C of previous external calibration. See page 234 for example calculations.