National Instruments Corporation 3-1 DAQ Hardware Overview Guide
3
LPM Devices
This chapter contains overview information on the PC-LPM-16 and
PC-LPM-16PnP.
LPM Device Analog Input
The LPM devices contain 16 multiplexed, single-ended analog input
channels numbered 0 through 15. The analog input channels are driven into
a 12-bit, self-calibrating ADC. The devices have no gains on the analog
input.
You can configure analog input (hardware jumpers on the PC-LPM-16 or
software configuration on the PC-LPM-16PnP) for four different input
ranges:
0 to +5 V (unipolar)
0 to +10 V (unipolar)
2.5 to +2.5 V (bipolar)
5 to +5 V (bipolar)
You can initiate A/D conversions through software or by applying active
low pulses to the EXTCONV* input on the I/O connector. On the
PC-LPM-16, a 16-word-deep FIFO memory on the device stores up to
16 A/D conversion results; on the PC-LPM-16PnP, a 256-word-deep FIFO
stores up to 256 A/D conversion results.
LPM Device Data Acquisition
The LPM devices can perform single-channel data acquisition and
multiple-channel scanned data acquisition. For single-channel data
acquisition, you select a single analog input channel. The device performs
a single A/D conversion on that channel every sample interval.
For multiple-channel scanned data acquisition, the device scans a sequence
of analog input channels. A sample interval indicates the time to elapse
between A/D conversions on each channel in the sequence. You need to
select only a single starting channel to select the sequence of channels to
scan. The device then scans the channels in consecutive order until
Seitenansicht 36